Patient Monitor 5 Function 12 inch_Meditech

20.000 EGP

  • Monitored Parameters: ECG, NIBP, SpO₂, Respiration, Temperature, Pulse Rate
  • ECG: 5-lead, ST segment & arrhythmia analysis, ±1 bpm / ±1% accuracy
  • Respiration: Impedance method, Adult 0–120 rpm, Pediatric/Neonatal 0–150 rpm, ±2 rpm accuracy
  • SpO₂: 1–99%, dual-wavelength LED, ±2% (70–99%) accuracy
  • NIBP: Oscillometric, Adult 40–240 mmHg, Pediatric 40–240 mmHg, Neonatal 10–150 mmHg, ±2 mmHg accuracy
  • Temperature: 0–50°C, ±0.1°C accuracy, 0.1°C resolution
  • Pulse Rate: 30–250 bpm, ±2 bpm or ±3% accuracy
  • Display: 12.1-inch color TFT, up to 8 waveforms
  • Battery: Built-in Lithium, up to 8 hours operation
  • Connectivity & Features: Trend storage 96 h, audible & visual alarms, LAN/Wireless, Central Monitoring support, protection vs electrosurgical interference
  • Optional: IBP, EtCO₂, Thermal Printer, OxyCRG

Patient Monitor 5 Function 12 inch_Meditech

20.000 EGP

In stock

Description

Standard Parameters

  • ECG (Electrocardiogram)
  • NIBP (Non-Invasive Blood Pressure)
  • SpO₂ (Oxygen Saturation)
  • RESP (Respiration Rate)
  • TEMP (Temperature)
  • PR (Pulse Rate)

 ECG

  • Lead Type: 5-lead
  • Lead Selection: I, II, III, aVR, aVL, aVF, V
  • Sweep Speed: 12.5 / 25 / 50 mm/s
  • Accuracy: ±1 bpm or ±1% (whichever is greater)
  • Protection: Up to 4000 VAC isolation against electrosurgical interference & defibrillation
  • Advanced Analysis: ST Segment & Arrhythmia
  • Alarms: Audible & visual with event recall

 Respiration (RESP)

  • Method: Impedance between RA–LL
  • Input Impedance: > 2.5 MΩ
  • Measurement Range: 0.3 – 5 Ω
  • Bandwidth: 0.3 – 2.5 Hz
  • Respiration Rate:
    • Adult: 0 – 120 rpm
    • Pediatric/Neonatal: 0 – 150 rpm
  • Resolution: 1 rpm
  • Accuracy: ±2 rpm
  • Apnea Alarm: 10 – 40 s

 SpO₂

  • Range: 1 – 99%
  • Resolution: 1%
  • Accuracy:
    • 70–99%: ±2%
    • 1–69%: ±5%
  • Method: Dual-wavelength LED

 NIBP

  • Technique: Oscillometric (during inflation)
  • Measurement Range:
    • Adult/Pediatric: 40 – 240 mmHg
    • Neonatal: 10 – 150 mmHg
  • Max Cuff Pressure:
    • Adult: 300 mmHg
    • Pediatric: 240 mmHg
    • Neonatal: 150 mmHg
  • Resolution: 1 mmHg
  • Accuracy: ±2 mmHg or ±2%
  • Measurement Modes: Manual / Automatic / STAT
  • Automatic Intervals: 2, 3, 5, 10, 15, 30 min / 1, 2, 4 h
  • STAT Mode: Continuous measurement for 5 min

 Temperature (TEMP)

  • Range: 0 – 50°C
  • Resolution: 0.1°C
  • Accuracy: ±0.1°C
  • Update Interval: ~1 second

 Pulse Rate (PR)

  • Range: 30 – 250 bpm
  • Accuracy: ±2 bpm or ±3%

 Printer (Optional)

  • Type: Thermal Printer
  • Paper Speed: 6.25 / 12.5 / 25 mm/s

 Power & Battery

  • Voltage: 110 – 240 V AC, 50 Hz
  • Power Consumption: ~8 W
  • Battery: Built-in Lithium
  • Operating Time: Up to 8 hours
  • Charging Time: ~4 hours

 Display & Interface

  • Screen Size: 12.1-inch color TFT
  • Resolution: 800 × 600 pixels
  • Waveforms Display: Up to 8 waveforms
  • Controls: 9 membrane keys + rotary knob (push & rotate)

 Additional Features

  • Trend data storage up to 96 hours
  • ST segment & arrhythmia analysis
  • Audible & visual alarms
  • LAN / Wireless connectivity
  • Central Monitoring System support (up to 64 monitors)
  • Protection against electrosurgical interference & defibrillation

 Optional Configurations

  • IBP (Invasive Blood Pressure)
  • EtCO₂ (End-tidal CO₂)
  • Thermal Printer
  • OxyCRG

Reviews

There are no reviews yet.


Be the first to review “Patient Monitor 5 Function 12 inch_Meditech”

General Inquiries

There are no inquiries yet.

Related Products